Hot Sale |
Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
Drug concentration calculation and titration table functions.
Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
With standard interface, big character interface and view bed, convenient to observe.
Display 7-lead ECG waveform on one screen.
Waveform, parameter color and location can be set optionally.
Fanless design,quiet,energy-saving and clean,which reduces the possibility of cross-infection.
7'' TFT color LCD,multi-language interface(English,Chinese,Russian,Kazakhstan, Czech, Slovak).
Adopt digital SpO2 technology, anti-motion and anti-ambient light interference,and measurement can be performed under the circumstance of weak filling.
NIBP measurement mode:Manual/AUTO/Continuous,storage for 4800-group NIBP data.
All-round monitor for adult, pediatric and neonate.
Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
This machine can monitor such parameters as ECG, RESP, SpO2, PR, NIBP and TEMP, etc. It integrates parameter measurement module, display and recorder in one device to form a compact and portable equipment.
Parameter:
ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: single-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
2, PVC, TACHY, BRADY, MISSED, PNP, PNC
Pacemaker: yes
RESP
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150 rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 12.5 mm/s, 25 mm/s
NIBP
Method: Oscillometry
Mode: Manual/AUTO/Continuous
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480 minutes
Measurement period in Continuous mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297±3 mmHg
Pediatric mode: 240±3 mmHg
Neonatal mode: 147±3 mmHg
SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified
PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater
TEMP
Channel: single-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1 ℃
Accuracy: ±0.1 ℃
Power supply
Adapter: 9 (V) DC
AC: 100 ~ 240 (V), 50/60 (Hz)
Maximal input power: 150VA
Safety classification: Class I, type CF defibrillation-proof applied part
Package List:
1 × Patient monitor
1 ×Adult fingertip SpO2 probe
1 ×SpO2 probe extension cable
1 ×Adult NIBP cuff
1 ×NIBP extension tube
1 ×ECG lead cable
1 ×ECG electrode
1 ×Temperature probe
1 ×Power cord
1 ×Power adapter
1 ×User Manual
FDA declaration : Statement:The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser.If the item is subject to FDA regulation, we will verify your status as an authorized purchaser of this item before shipping of the item.(Seller Name: Sara;Country: China;City: Beijing;TEL.:86-15210656055). This item has been cleaned and treated according to the manufacturer's instructions. The Fingertip Pulse Oximeter is certified with the US FDA 510K No. K070371, the CE & TUV of Europe and it is on the Australian Register of Therapeutic Goods (ARTG) with the code 136606. The Powered Surgical Instrument / Speed 808 System is certified with the US FDA 510(k) Number:K132989 The Powered Surgical Instrument / Hair Remove Device is certified with the US FDA 510(k) Number:K180353 The Powered Surgical Instrument / Hair Remove System is certified with the US FDA 510(k) Number:K141973 massager, vacuum, light induced heating / Slimming Treatment Device is certified with the US FDA 510(k) Number:K161892New Arrival |